Establishing the presence of cochlear implant related artifact during sound-field recording of the auditory steady state response: A comparison between normal hearing adults, cochlear implant recipients, and a cochlear implanted human cadaver

Journal of the Acoustical Society of America(2017)

引用 0|浏览14
暂无评分
摘要
The Auditory Steady State Response (ASSR) is an objective technique permitting dichotic, frequency-specific hearing threshold estimation. Electrical ASSR (eASSR) has the potential for auditory assessment in cochlear implant (CI) users through direct CI electrode stimulation. However, sound-field ASSR (sASSR) provides a more natural listening environment for CI users. Electrophysiological recordings could be affected by CI related artifact. The present study investigated the possibility of CI-artifact during sASSR recording in 10 CI subjects at 0.5, 1, 2 and 4 kHz. Results were compared with 15 NH controls. We investigated the mean threshold differences between the sASSR thresholds and the behavioral thresholds for each group at the four frequencies. Correlation between the sASSR thresholds and behavioral thresholds were computed. There were significant differences between the NH and CI groups across the four frequencies indicating the possibility of a CI-related artifact. In order to confirm the presence ...
更多
查看译文
关键词
Auditory Processing,Speech Perception,Cochlear Implants
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要