•Authors developed the new parameter of current coil that they named PCE.•There are proposed the calibration method and the calibration procedure.•There is proposed the automatic system to determine ACE and PCE of current coils.•Power error due to the influence of PCE is higher than accuracy class of meters.•It is obvious that influence of the PCE cannot be omitted.
更多
查看译文
关键词
Power clamp meter,Current clamp meter,Current coil,Coil effect,Clamp/coil interaction,Power calibrator,Calibration procedure,Measurement uncertainty