Shim Coil Set for an Open Biplanar MRI System Using an Inverse Boundary Element Method

IEEE Transactions on Applied Superconductivity(2016)

引用 8|浏览10
暂无评分
摘要
This paper proposes a numerical method with an inverse boundary element method (IBEM) to design a set of shim coils for an open bipanar superconducting magnetic resonance imaging (MRI) system. For the traditional cylindrical MRI system, claustrophobics will suffer the palpitations and anxiety attacks when they are enclosed in a small area during the scanning process. However, such bipanar structur...
更多
查看译文
关键词
Magnetic fields,Magnetic resonance imaging,Current density,Wires,Superconducting magnets,Harmonic analysis,Surface treatment