Time Dependent DD Neutrons Measurement Using a Single Crystal Chemical Vapor Deposition Diamond Detector on EAST

PLASMA SCIENCE & TECHNOLOGY(2016)

引用 7|浏览34
暂无评分
摘要
A single crystal chemical vapor deposition (scCVD) diamond detector has been successfully employed for neutron measurements in the EAST (Experimental Advanced Superconducting Tokamak) plasmas. The scCVD diamond detector coated with a 5 mu m (LiF)-Li-6 (95% Li-6 enriched) layer was placed inside a polyethylene moderator to enhance the detection efficiency. The time-dependent neutron emission from deuteron plasmas during neutral beam injection (NBI) heating was obtained. The measured results are compared with that of fission chamber detectors, which always act as standard neutron flux monitors. The scCVD diamond detector exhibits good reliability, stability and the capability to withstand harsh radiation environments despite its low detection efficiency due to the small active volume.
更多
查看译文
关键词
scCVD diamond detector,neutron measurement,EAST
AI 理解论文