Validation of the Microlife BP A200 Comfort and W2 Slim automated blood pressure monitors in a general adult population according to the European Society of Hypertension and the ANSI/AAMI/ISO 81060-2: 2013 protocols.

BLOOD PRESSURE MONITORING(2016)

引用 7|浏览1
暂无评分
摘要
ObjectiveThis study aimed to determine the accuracy of the Microlife BP A200 Comfort and W2 Slim automated blood pressure monitors according to the European Society of Hypertension International Protocol revision 2010 and the ANSI/AAMI/ISO 81060-2:2013 protocols.MethodsThe devices were assessed on 33 participants according to the European Society of Hypertension requirements and were then tested on 85 participants according to the ANSI/AAMI/ISO 81060-2:2013 criteria. Procedures and data analysis were carried out following protocol guidelines precisely.ResultsThe Microlife BP A200 Comfort and W2 Slim devices passed the criteria of the European Society of Hypertension International Protocol revision 2010 for both systolic blood pressure and diastolic blood pressure. The devices also fulfilled the ANSI/AAMI/ISO 81060-2:2013 criteria, with mean differences of SBP and DPB between the devices and observers of 0.385.12 and 0.28 +/- 4.29mmHg for the BP A200 Comfort and 1.01 +/- 6.80 and 0.34 +/- 5.62mmHg for the W2 Slim, respectively.ConclusionThe Microlife BP A200 Comfort and W2 Slim automated blood pressure monitors fulfilled the European Society of Hypertension revision 2010 and the ANSI/AAMI/ISO 81060-2:2013 protocols, and can be recommended for self-measurement in the general population.
更多
查看译文
关键词
blood pressure,oscillometry,self-measurement,validation