[Consideration on SNR in Synthetic MRI].

Akihito Ikeda,Kohki Yoshikawa

Nihon Hoshasen Gijutsu Gakkai zasshi(2019)

引用 2|浏览2
暂无评分
摘要
Synthetic magnetic resonance imaging (MRI) can create different contrast weighted images by quantifying the T, T, and proton density values of the subjects from a single series of scan data. It has not been clarified how the signal to noise ratio (SNR) of the synthesized image varies depending on imaging parameters. We investigated the change of SNR in synthesized MR images by the experiment using self-made phantom. The SNR ratio of synthesized image by synthetic MRI showed the same tendency as the theoretical values due to parameter change in N, N, slice thickness, number of excitations. However, as for BW, the SNR ratio tended to be different from the theoretical values in some cases. In addition, it was suggested that the SNR of the composite image has relevance to the quantitative accuracy of the T, T, and proton density values. We thought that this is due to the image acquisition process by synthetic MRI.
更多
查看译文
关键词
coefficient of variation (CV),magnetic resonance imaging (MRI),quantification of relaxation times and proton density by multi-echo acquisition of saturation-recovery using turbo spin-echo readout (QRAPMASTER),signal to noise ratio (SNR),synthetic MRI