Lack Of Harmonization In High Fluorescent Cell Automated Counts With Body Fluids Mode In Ascitic, Pleural, Synovial, And Cerebrospinal Fluids

INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY(2019)

引用 11|浏览3
暂无评分
摘要
IntroductionCellular analysis in body fluids (BFs) provides important diagnostic information in various pathological settings. This study was hence aimed at comparing automated cell count obtained with Mindray BC-6800 (BC-BF) vs Sysmex XN-series (XN-BF) and evaluating other quantitative and qualitative information provided by these analyzers in ascitic (AF), pleural (PF), synovial (SF), and cerebrospinal (CSF) fluids.MethodsThree hundred and fifty-one samples (99 AFs, 45 PFs, 75 SFs, and 132 CSFs) were analyzed in parallel with BC-BF, XN-BF, and optical microscopy (OM). The study also included the assessment of diagnostic agreement among BC-BF, XN-BF, and OM.ResultsThe comparison of BC-BF vs XN-BF yielded slopes of Passing and Bablok regression always comprised between 0.9 and 1.0 except for EO-BF and HF-BF, whilst the intercepts ranged from -0.4 for MN-BF and 12.0 for PMN-BF. The bias was comprised between -103.3% and 67.1% for HF-BF and EO-BF, respectively. A significant bias was found for TC-BF, WBC-BF, HF-BF (negative bias) and for PMN-BF and EO-BF (positive bias). The agreement (Cohen's kappa) between XN-BF and BC-BF was always 0.7, ranging between 0.87 in CSFs and 0.94 in AFs, and that with OM was similar (ie, 0.85 and 0.96).ConclusionThe cytometric analysis of BF samples using BC-BF and XN-BF is clinically favorable when appropriately combined with OM. Quantitative and qualitative parameters displayed optimal agreement, whilst instrument-specific cut-offs should be defined and implemented for HF-BF and EO-BF. Further efforts should be made for achieving better harmonization in cytometric analysis of BF samples.
更多
查看译文
关键词
automated cell count, BC-6800, body fluid, optical microscopy, XN
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要