Clinical evaluation of the repeatability of ocular aberrometry obtained with a new pyramid wavefront sensor.

EUROPEAN JOURNAL OF OPHTHALMOLOGY(2019)

引用 15|浏览6
暂无评分
摘要
Purpose: To evaluate the intrasubject repeatability of the ocular aberrometry obtained with a new ocular pyramidal aberrometer technology in a sample of normal eyes. Methods: A total of 53 healthy eyes of 53 subjects with ages ranging from 18 to 45 years were included in this study. In all cases, three consecutive acquisitions were obtained. Intrasubject repeatability of the measurements with a pyramidal aberrometer was calculated. Intrasubject repeatability for 4.0- and 6.0-mm pupils was evaluated within the subject standard deviation (S-w) and intraclass correlation coefficient. Results: Low values of the S-w and intraclass correlation coefficient outcomes close to 1 were observed for the sphere and cylinder at 3.0-mm pupil size. Most low S-w and intraclass correlation coefficient values close to 1 were observed for total, low-order aberrations and higher-order aberrations root mean square and for each Zernike coefficient analysis (intraclass correlation coefficient > 0.798) at 4.0-mm pupil size, with more limited outcomes for the aberrometric coefficient of Z(4, 4) with an intraclass correlation coefficient of 0.683. For a 6.0 mm pupil diameter, low S-w and intraclass correlation coefficient values close to 1 were observed for all aberrometric parameters or Zernike coefficients analyzed (intraclass correlation coefficient > 0.850). Conclusion: The new pyramidal aberrometer Osiris provides repeatable and consistent measurements of ocular aberrometry measurements in normal eyes.
更多
查看译文
关键词
Ocular aberrometry,pyramidal aberrometer,intrasubject repeatability
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要