Comparison of dosimetric properties among four commercial multi-detector computed tomography scanners

Physica Medica(2017)

引用 3|浏览9
暂无评分
摘要
•TG-111 reported new CT dosimetry using Farmer chamber in cylindrical water phantom.•We compare absorbed dose among four MDCT based on TG-111.•Our method made it possible to directly measure the absorbed dose with small uncertainty.•Absorbed dose was calculated using 60Co absorbed dose-to-water calibration factor.•Four MDCT scanners were modeled in Monte Carlo to correspond with measured values.
更多
查看译文
关键词
Multi-detector CT,Monte Carlo technique,Ionization chamber measurement,Absorbed dose