Personnel real time dosimetry in interventional radiology

L. Servoli,L. Bissi,S. Fabiani,D. Magalotti,P. Placidi,A. Scorzoni, A. Calandra, R. Cicioni, S. Chiocchini, A.C. Dipilato, N. Forini,M. Paolucci,R. Di Lorenzo, F.P. Cappotto, M. Scarpignato, A. Maselli, A. Pentiricci

Physica Medica(2016)

引用 8|浏览33
暂无评分
摘要
Interventional radiology and hemodynamic procedures have rapidly grown in number in the past decade, increasing the importance of personnel dosimetry not only for patients but also for medical staff. The optimization of the absorbed dose during operations is one of the goals that fostered the development of real-time dosimetric systems. Indeed, introducing proper procedure optimization, like correlating dose rate measurements with medical staff position inside the operating room, the absorbed dose could be reduced. Real-time dose measurements would greatly facilitate this task through real-time monitoring and automatic data recording. Besides real-time dose monitoring could allow automatic data recording. In this work, we will describe the calibration and validation of a wireless real-time prototype dosimeter based on a new sensor device (CMOS imager). The validation measurement campaign in clinical conditions has demonstrated the prototype capability of measuring dose-rates with a frequency in the range of few Hz, and an uncertainty smaller than 10%.
更多
查看译文
关键词
CMOS imager,Interventional radiology,Dosimetry