Evaluation of the V8 E-Class, a Novel Automated Capillary Isoelectric Focusing Instrument for Hemoglobinopathy Screening.

AMERICAN JOURNAL OF CLINICAL PATHOLOGY(2016)

引用 4|浏览8
暂无评分
摘要
Objectives: We evaluated the performance of a novel capillary isoelectric focusing (CIEF) application for hemoglobinopathy screening on the recently introduced V8 E-Class platform. Methods: Analytical performance of the V8 E-Class was evaluated and included assessment of hemoglobin A(2) (HbA(2)) imprecision; linearity for HbA(2), fetal hemoglobin (HbF), and sickle hemoglobin (HbS); and carryover for HbS. Furthermore, a method comparison with the Minicap Flex Piercing (Sebia, Lisses, France), the Variant Classic (Bio-Rad Laboratories, Hercules, CA), and the G8 (Tosoh Europe, Amsterdam, the Netherlands) was done to assess analytical and clinical concordance. Results: Total HbA(2) imprecision was 3.26% and 3.14% for normal and elevated HbA(2) controls and 5.16% and 3.58% for a normal and a heterozygous HbS patient sample, respectively. HbA(2), HbF, and HbS showed acceptable linearity, and no carryover was observed. The method comparison showed good analytical concordance (r > 0.95) except for a homozygous HbS subset (r = 0.532-0.704). A comparable phenomenon was seen for the clinical concordance with good agreement in samples without variants (weighted kappa > 0.80) but poorer agreement in HbS samples (kappa < 0.30). Conclusions: Good analytical performance was demonstrated for this novel CIEF application for hemoglobinopathy screening. Method comparison showed generally good correlation but highlights the need for standardization. Finally, software optimization could further add to its use for routine hemoglobinopathy screening.
更多
查看译文
关键词
Capillary isoelectric focusing,Hemoglobinopathy screening,Helena Biosciences V8 E-Class,Cation-exchange HPLC,Capillary zone electrophoresis
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要