Validation Of Workflow Changes, Phage Concentration And Reformatted Detection Threshold For The Sample6 Detect/L (Tm) Test: Level 3 Modification

JOURNAL OF AOAC INTERNATIONAL(2018)

引用 2|浏览18
暂无评分
摘要
The AOAC Research Institute Performance Tested Methods(SM) Program certified Sample6 DETECT/L (TM) in April 2014 (Certification No. 041401) for the detection of Listeria species (L. monocytogenes, L. innocua, L. ivanovii, L. seeligeri, L. marthii, L. welshimeri) on stainless steel environmental surfaces. A modification was approved in January 2016, increasing the concentration of sanitizer-neutralizing reagents in detection reagents, increasing the number of phage in the detection solution, and increasing the sample test volume. Moreover, changes to reduce the number of negative controls and add compatibility with polyurethane sponges were also approved. In this modification, to ensure that DETECT/L continues to meet performance expectations, Sample6 evaluated workflow changes to enhance sensitivity and the ease-of-use of the assay. Changes to the phage concentration and detection threshold, plus the inclusion of a confirmation step (DETECT Check), were validated to obtain better accuracy and optimize assay performance. Inclusivity, exclusivity, and robustness testing were conducted by Sample6 to evaluate the changes. A third-party laboratory compared the DETECT/L assay and the U.S. Department of Agriculture reference method in a stainless steel environmental surface matrix study. The data presented in this report demonstrate that the changes proposed to the DETECT/L assay meet or exceed the performance in the current configuration.
更多
查看译文
关键词
sample6 detection,phage concentration,reformatted detection threshold,workflow changes,test
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要