Challenging the six-hour recommendation for reprocessing sterilizable medical equipment.

Journal of Hospital Infection(2019)

引用 7|浏览7
暂无评分
摘要
No association was identified between residual protein and holding time, nor between incidence of corrosion and holding time. The study thereby challenges the relevance of upholding the recommendation of a maximum wait of 6 h prior to reprocessing. The findings will potentially have an impact on the organization of reprocessing of surgical instruments in Denmark and internationally.
更多
查看译文
关键词
Reprocessing,Surgical instruments,Holding time,Protein residue,Corrosion