Evaluating The Analytical Performance Of Four New Coagulation Assays For The Measurement Of Fibrinogen, D-Dimer And Thrombin Time

INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY(2018)

引用 9|浏览15
暂无评分
摘要
Introduction New laboratory methods to measure haemostatic function require careful assessment before routine use. We evaluated the analytical performance of four new coagulation assays for the measurement of fibrinogen by Clauss assay, prothrombin time-derived fibrinogen, thrombin time and D-dimer levels. Methods Results The four assays were evaluated on the cobas t 711 and cobas t 511 analysers at four centres in Europe. Analytical performance and method comparisons with other commercially available assays were performed according to Clinical and Laboratory Standards Institute guidelines (EP09-A3, EP05-A3) using residual anonymized human sodium citrate (3.2% [0.109M]) plasma samples. Lot-to-lot variability and the equivalency of each assay on the cobas t 711 and cobas t 511 analysers were also assessed. Overall, coefficients of variance were <= 4.1% and <= 8.6% for within-run precision and total reproducibility, respectively. Method comparison experiments showed good or acceptable agreement for each assay compared with their respective comparator method, and equivalency was demonstrated for the two cobas t platforms (Pearson's correlation coefficient >= 0.991). A high level of consistency was observed between lots for all four assays (Pearson's correlation coefficient >= 0.994). Conclusion This multicentre study demonstrates excellent analytical performance for four new coagulation assays on the cobas t 711 and cobas t 511 analysers.
更多
查看译文
关键词
cobas t 511, cobas t 711, D-dimer, fibrinogen, thrombin time
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要