Biomechanics of common fixation devices for first tarsometatarsal joint fusion-a comparative study with synthetic bones.

Journal of orthopaedic surgery and research(2018)

引用 21|浏览26
暂无评分
摘要
The intramedullary device demonstrated the highest initial compression force of the three tested implants. The plantar locking plate showed the best overall stability during weight-bearing simulation. Further clinical research is necessary to analyse if the intramedullary fixation device needs a longer period of non-weight-bearing to reach a better non-union rate compared to the plantar locking plate.
更多
查看译文
关键词
Angle stable,Hallux valgus,Intramedullary fixation device,Lapidus fusion,Locking plate