Dual-energy CT and ceramic or titanium prostheses material reduce CT artifacts and provide superior image quality of total knee arthroplasty

Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA(2018)

引用 18|浏览17
暂无评分
摘要
Purpose To evaluate the influence of different scan parameters for single-energy CT and dual-energy CT, as well as the impact of different material used in a TKA prosthesis on image quality and the extent of metal artifacts. Methods Eight pairs of TKA prostheses from different vendors were examined in a phantom set-up. Each pair consisted of a conventional CoCr prosthesis and the corresponding anti-allergic prosthesis (full titanium, ceramic, or ceramic-coated) from the same vendor. Nine different (seven dual-energy CT and two single-energy CT) scan protocols with different characteristics were used to determine the most suitable CT protocol for TKA imaging. Quantitative image analysis included assessment of blooming artifacts (metal implants appear thicker on CT than they are, given as virtual growth in mm in this paper) and streak artifacts (thick dark lines around metal). Qualitative image analysis was used to investigate the bone–prosthesis interface. Results The full titanium prosthesis and full ceramic knee showed significantly fewer blooming artifacts compared to the standard CoCr prosthesis (mean virtual growth 0.6–2.2 mm compared to 2.9–4.6 mm, p < 0.001). Dual-energy CT protocols showed less blooming (range 3.3–3.8 mm) compared to single-energy protocols (4.6–5.5 mm). The full titanium and full ceramic prostheses showed significantly fewer streak artifacts (mean standard deviation 77–86 Hounsfield unit (HU)) compared to the standard CoCr prosthesis (277–334 HU, p < 0.001). All dual-energy CT protocols had fewer metal streak artifacts (215–296 HU compared to single-energy CT protocols (392–497 HU)). Full titanium and ceramic prostheses were ranked superior with regard to the image quality at the bone/prosthesis interface compared to a standard CoCr prosthesis, and all dual-energy CT protocols were ranked better than single-energy protocols. Conclusions Dual-energy CT and ceramic or titanium prostheses reduce CT artifacts and provide superior image quality of total knee arthroplasty at the bone/prosthesis interface. These findings support the use of dual-energy CT as a solid imaging base for clinical decision-making and the use of full-titanium or ceramic prostheses to allow for better CT visualization of the bone–prosthesis interface.
更多
查看译文
关键词
Total knee arthroplasty,Single-energy CT,Dual-energy CT,Blooming artifacts,Streak artifacts,Mono-energetic imaging
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要