Integrated catheter for simultaneous radio frequency ablation and optoacoustic monitoring of lesion progression.

OPTICS LETTERS(2018)

引用 25|浏览10
暂无评分
摘要
Radio frequency (RF) catheter ablation is commonly used to eliminate dysfunctional cardiac tissue by heating via an alternating current. Clinical outcomes are highly dependent on careful anatomical guidance, electrophysiological mapping, and careful RF power titration during the procedure. Yet, current treatments rely mainly on the expertise of the surgeon to assess lesion formation, causing large variabilities in the success rate. We present an integrated catheter design suitable for simultaneous RF ablation and real-time optoacoustic monitoring of the forming lesion. The catheter design utilizes copper-coated multimode light guides capable of delivering both ablation current and near-infrared pulsed-laser illumination to the target tissue. The generated optoacoustic responses were used to visualize the ablation lesion formation in an ex-vivo bovine heart specimen in 3D. The presented catheter design enables the monitoring of ablation lesions with high spatiotemporal resolution while the overall therapy-monitoring approach remains compatible with commercially available catheter designs. (C) 2018 Optical Society of America