Comparison of 3 He and 129 Xe MRI for evaluation of lung microstructure and ventilation at 1.5T.

JOURNAL OF MAGNETIC RESONANCE IMAGING(2018)

引用 67|浏览8
暂无评分
摘要
BACKGROUND:To support translational lung MRI research with hyperpolarized 129 Xe gas, comprehensive evaluation of derived quantitative lung function measures against established measures from 3 He MRI is required. Few comparative studies have been performed to date, only at 3T, and multisession repeatability of 129 Xe functional metrics have not been reported. PURPOSE/HYPOTHESIS:To compare hyperpolarized 129 Xe and 3 He MRI-derived quantitative metrics of lung ventilation and microstructure, and their repeatability, at 1.5T. STUDY TYPE:Retrospective. POPULATION:Fourteen healthy nonsmokers (HN), five exsmokers (ES), five patients with chronic obstructive pulmonary disease (COPD), and 16 patients with nonsmall-cell lung cancer (NSCLC). FIELD STRENGTH/SEQUENCE:1.5T. NSCLC, COPD patients and selected HN subjects underwent 3D balanced steady-state free-precession lung ventilation MRI using both 3 He and 129 Xe. Selected HN, all ES, and COPD patients underwent 2D multislice spoiled gradient-echo diffusion-weighted lung MRI using both hyperpolarized gas nuclei. ASSESSMENT:Ventilated volume percentages (VV%) and mean apparent diffusion coefficients (ADC) were derived from imaging. COPD patients performed the whole MR protocol in four separate scan sessions to assess repeatability. Same-day pulmonary function tests were performed. STATISTICAL TESTS:Intermetric correlations: Spearman's coefficient. Intergroup/internuclei differences: analysis of variance / Wilcoxon's signed rank. Repeatability: coefficient of variation (CV), intraclass correlation (ICC) coefficient. RESULTS:A significant positive correlation between 3 He and 129 Xe VV% was observed (r = 0.860, P < 0.001). VV% was larger for 3 He than 129 Xe (P = 0.001); average bias, 8.79%. A strong correlation between mean 3 He and 129 Xe ADC was obtained (r = 0.922, P < 0.001). MR parameters exhibited good correlations with pulmonary function tests. In COPD patients, mean CV of 3 He and 129 Xe VV% was 4.08% and 13.01%, respectively, with ICC coefficients of 0.541 (P = 0.061) and 0.458 (P = 0.095). Mean 3 He and 129 Xe ADC values were highly repeatable (mean CV: 2.98%, 2.77%, respectively; ICC: 0.995, P < 0.001; 0.936, P < 0.001). DATA CONCLUSION: 129 Xe lung MRI provides near-equivalent information to 3 He for quantitative lung ventilation and microstructural MRI at 1.5T. LEVEL OF EVIDENCE:3 Technical Efficacy Stage 2 J. Magn. Reson. Imaging 2018.
更多
查看译文
关键词
chronic obstructive pulmonary disease,helium-3,hyperpolarized gas,lung MRI,lung cancer,repeatability,xenon-129
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要