Transcatheter Valve-In-Valve Overexpansion For Treating A Large Dysfunctional Tricuspid Bioprosthesis

INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY(2018)

引用 2|浏览10
暂无评分
摘要
Transcatheter valve-in-valve implantation within dysfunctional surgical bioprosthesis has become an alternative to redo open-heart surgery. However, suitability for valve-in-valve implantation in the tricuspid position is often limited by large surgical valve sizes. We report a case of a transcatheter tricuspid valve-in-valve implantation with a 29-mm balloon-expandable prosthesis within a 33-mm failed bioprosthesis (exceeding manufacturer's sizing recommendations). Overexpansion of a 29-mm SAPIEN XT valve by 1-ml overfilling of the deployment balloon was successfully performed, with no valve dislocation/embolization or residual tricuspid regurgitation. This case illustrates the feasibility of tricuspid valve-in-valve procedures in selected patients with large failed tricuspid bioprostheses.
更多
查看译文
关键词
Transcatheter valve replacement, Tricuspid valve, Valve in valve, Overfilling