Feasibility of intraoperative detection of sentinel lymph nodes with 89-zirconium-labelled nanocolloidal albumin PET-CT and a handheld high-energy gamma probe

EJNMMI research(2018)

引用 29|浏览26
暂无评分
摘要
Background PET/CT lymphoscintigraphy using 89 Zr-nanocolloidal albumin has the potential to improve the preoperative identification of sentinel lymph nodes (SLNs), especially if located in the near proximity of the primary tumour. This study aims to demonstrate the feasibility of PET/CT lymphoscintigraphy followed by intraoperative detection of 89 Zr-nanocolloidal albumin containing SLNs with the use of a handheld high-energy gamma probe. Methods PET/CT lymphoscintigraphy was performed after peritumoural injection of 89 Zr-nanocolloidal albumin in five patients with oral cavity carcinoma planned for surgical resection. SLN biopsy procedure was performed 18 h later. SLNs were detected using detailed information of PET/CT and the high-energy gamma probe. Results In all patients, SLNs were identified on PET/CT lymphoscintigraphy. Intraoperative detection using the high-energy gamma probe was possible in 10 of 13 SLNs, at a short distance from the SLN. Conclusions This study demonstrates that intraoperative detection of SLNs containing 89 Zr-nanocolloidal albumin using a handheld high-energy gamma probe is feasible, but its clinical use and sensitivity seem to be limited. Trial registration CCMO NL37222.092.11
更多
查看译文
关键词
89Zr-nanocolloidal albumin,Head and neck cancer,High-energy gamma probe,PET/CT lymphoscintigraphy,Sentinel lymph node