Comparison of maximal Rubidium-82 activities for myocardial blood flow quantification between digital and conventional PET systems

Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology(2018)

引用 17|浏览10
暂无评分
摘要
Background PET-based myocardial blood flow (MBF) quantification can be inaccurate when using high tracer activities. Our aim was to derive the maximal Rubidium-82 activity for MBF assessment using a new digital PET system and compare the results with conventional analog systems. Methods 1.8 GBq Rubidium-82 was injected into the cardiac insert of an anthropomorphic torso phantom. Data were acquired for 10 min using an Ingenuity TF (Philips Healthcare), Discovery 690 (D690, GE Healthcare), and digital PET prototype system (Philips Healthcare). The dynamic ranges, defined as the maximal measured activity in the reconstructed images deviating < 10% from the true present activity, were determined in all scans. Results The dynamic ranges were 312 MBq for Ingenuity TF, 650 MBq for D690, and 654 MBq for digital PET prototype. Conclusions The maximal Rb-82 activity for MBF assessment using digital PET prototype is higher than that for its analog counterpart (Ingenuity TF), but seems comparable to the D690.
更多
查看译文
关键词
Digital PET,direct photon counting,dynamic range,cardiac positron emission tomography,rubidium-82