Retrieval and radiographic analysis of the Contour antiprotrusio cage.

HIP INTERNATIONAL(2017)

引用 11|浏览11
暂无评分
摘要
Introduction: Acetabular reconstruction in the setting of severe bone loss or pelvic discontinuity remains a challenging problem. Multiple methods of treatment have been described including antiprotrusio cages (APCs). The objective of this study is to combine biomechanical analysis of retrieved APCs with radiographic and clinical data to determine which factors influence or predict APC failure. Methods: 41 APCs were identified. Sequential radiographs were examined for cage and polyethylene cup abduction angles, change in centre of rotation, screw placement, progression of cage failure, and failure mechanism. Cages were manually examined for gross macroscopic findings, breakage, and the location of breakage. Highresolution microscopy was used for further analysis. Results: 24 cages were included in the analysis. Mean age of patients was 64.5 years (range 43-85 years); average length of implantation was 42.5 months (range 3-108 months). Average cage abduction angles were 56 degrees; abduction for the cemented polyethylene cup was 44 degrees. 14 of 24 cages were broken; 10 were intact. Of the broken cages, 10/14 broke through a screw hole in the ischial flange or just superior to the ischial flange. In the intact group, 6/10 failed due to pullout of the ischial screws. Discussion: All cages had superior and lateralised centres of rotation. The majority of cages failed due to breakage or pullout at the ischial flange. Pelvic discontinuity was a large risk factor for a broken cage. Future design and technique modifications may result in superior outcomes in these complex acetabular reconstructions.
更多
查看译文
关键词
Acetabular reconstruction,Antiprotrusio,Cage,Pelvic discontinuity,Revision total hip arthroplasty
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要