A Multi-site In-depth Evaluation of the Quanterix Simoa from a User’s Perspective

The AAPS journal(2017)

引用 16|浏览33
暂无评分
摘要
An in-depth evaluation of the Quanterix© Simoa™ platform was undertaken by scientists from the AAPS Emerging Technologies Focus Group to determine the overall performance of the technology as well as provide guidance to future users. In order to test the platform in a non-GLP bioanalytical setting, a cross-site evaluation of the Quanterix IL-6 biomarker kit was performed. Parameters tested during this evaluation included sensitivity, accuracy and precision, and parallelism in human serum from normal individuals. The results demonstrated improved sensitivity compared to the claimed sensitivity of other commercially available IL-6 kits and showed excellent site-to-site reproducibility. Observed issues included difficulties with system reliability and a lack of parallelism and specificity in a subset of samples. Overall, these results demonstrate that while there are challenges to the Simoa platform this technology offers automation capabilities and excellent sensitivity that enhance bioanalysis especially of low-abundance analytes.
更多
查看译文
关键词
automation,IL-6,ligand binding assay,Simoa,ultrasensitive technology