Validation of the Australian diagnostic reference levels for paediatric multi detector computed tomography: a comparison of RANZCR QUDI data and subsequent NDRLS data from 2012 to 2015

Australasian Physical & Engineering Sciences in Medicine(2016)

引用 1|浏览1
暂无评分
摘要
The national diagnostic reference level service (NDRLS), was launched in 2011, however no paediatric data were submitted during the first calendar year of operation. As such, Australian national diagnostic reference levels (DRLs), for paediatric multi detector computed tomography (MDCT), were established using data obtained from a Royal Australian and New Zealand College of Radiologists (RANZCR), Quality Use of Diagnostic Imaging (QUDI), study. Paediatric data were submitted to the NDRLS in 2012 through 2015. An analysis has been made of the NDRLS paediatric data using the same method as was used to analyse the QUDI data to establish the Australian national paediatric DRLs for MDCT. An analysis of the paediatric NDRLS data has also been made using the method used to calculate the Australian national adult DRLs for MDCT. A comparison between the QUDI data and subsequent NDRLS data shows the NDRLS data to be lower on average for the Head and AbdoPelvis protocol and similar for the chest protocol. Using an average of NDRLS data submitted between 2012 and 2015 implications for updated paediatric DRLS are considered.
更多
查看译文
关键词
Diagnostic reference level,Paediatric,Computed tomography,Dose length product,CTDI,vol