The efficiency calibration of radioxenon isotopes of inhomogeneous adsorption column for HPGe γ spectrometry
Journal of Radioanalytical and Nuclear Chemistry(2018)
摘要
In this paper, a new efficiency calibration scheme is proposed for the inhomogeneous adsorption column, in which the segment efficiencies modeled by the Monte Carlo toolkit Geant4 are weighted according to the fraction distributions of the adsorbed radioxenon based on the equilibrium adsorption theory. In order to verify the scheme, an experiment of efficiency calibration of HPGe detector by employing radon daughters was performed, and as a result the relative errors between the measured value and simulated value was less than ± 2.3%. This method developed in this paper is useful for the development of rapid test device of radioactive gas.
更多查看译文
关键词
Efficiency calibration,Radioxenon isotopes,Inhomogeneous adsorption column
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要