An implantable mixed-signal CMOS die for battery-powered in vivo blowfly neural recordings.

Microelectronics Journal(2018)

引用 7|浏览16
暂无评分
摘要
A mixed-signal die containing two differential input amplifiers, a multiplexer and a 50 KSPS, 10-bit SAR ADC, has been designed and fabricated in a 0.35 μm CMOS process for in vivo neural recording from freely moving blowflies where power supplied voltage drops quickly due to the space/weight limited insufficient capacity of the battery. The designed neural amplifier has a 66 + dB gain, 0.13 Hz-5.3 KHz bandwidth and 0.39% THD. A 20% power supply voltage drop causes only a 3% change in amplifier gain and 0.9-bit resolution degrading for SAR ADC while the on-chip data modulation reduces the chip size, rendering the designed chip suitable for battery-powered applications. The fabricated die occupies 1.1 mm2 while consuming 238 μW, being suitable for implantable neural recordings from insects as small as a blowfly for electrophysiological studies of their sensorimotor control mechanisms. The functionality of the die has been validated by recording the signals from identified interneurons in the blowfly visual system.
更多
查看译文
关键词
Implantable neural recording,Blowfly,CMOS neural amplifier,SAR ADC
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要