Application of double-focusing sector field ICP-MS for determination of ultratrace constituents in samples characterized by complex composition of the matrix.

Science of The Total Environment(2018)

引用 25|浏览7
暂无评分
摘要
The performance of double focusing, sector field mass spectrometry (ICP-SFMS) for determination of analytes, including technology critical elements (TCE), at ultra-trace levels in environmental and clinical matrices was critically evaluated. Different configurations of the ICP-SFMS introduction system as well as various sample preparations, pre-concentration and matrix separation methods were employed and compared. Factors affecting detection capabilities and accuracy of data produced (instrumental sensitivity, contamination risks, purity of reagents, spectral interferences, matrix effects, analyte recovery and losses) were discussed. Optimized matrix-specific methods were applied to a range of reference and control materials (riverine, brackish and seawaters; whole blood, serum and urine) as well as tap water and snow samples collected in the area of Luleå city, northern Sweden; brackish and seawater from the Laptev Sea; venous blood samples with a special emphasis on determination of Au, Ag, Ir, Os, Pd, Pt, Re, Rh, Ru, Sb and Te. Even though these low abundant elements are relatively under-documented, the results produced were compared with published data, where available.
更多
查看译文
关键词
ICP-SFMS,TCE,PGE,Natural waters,Snow,Clinical samples