Design Of A Compact Collimator And 3d Imaging System For A Scanning Beam Low-Energy Intraoperative Radiation Therapy System For Pancreatic Cancer
2017 39TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC)(2017)
摘要
Intraoperative radiation therapy (IORT) involves delivering high doses of radiation directly to tumors while sparing healthy tissues in a surgical setting. Current IORT systems are limited in their lack of image guidance and variable needs for shielded operating rooms. They also lack the capability to deliver non-uniform therapeutic radiation to irregular shaped clinical targets. We developed a scanning beam IORT system (SBIORT) to overcome these limitations. SBIORT consists of a low energy x-ray source, a custom compact dynamic x-ray collimator system, a robotic arm, and a 3D surface imaging module. Here we describe the design and validation of the compact dynamic x-ray collimator system and the 3D surface imaging module for use in SBIORT. The proposed collimator can achieve a leaf position accuracy of +/- 0.25 mm (95% confidence interval). Phantom studies indicated the 3D surface-imaging module has an accuracy of 1.0 +/- 0.6 mm with ability to obtain high resolution surface image within 5 seconds. SBIORT is a novel approach to deliver conformal intensity-modulated intraoperative radiation therapy.
更多查看译文
关键词
Humans,Imaging, Three-Dimensional,Pancreatic Neoplasms,Phantoms, Imaging,Radiotherapy Dosage,Radiotherapy, Intensity-Modulated