An EWOD-based micro diluter with high flexibility on dilution ratio

Microsystem Technologies(2016)

引用 13|浏览5
暂无评分
摘要
Dilution is a common procedure in biological and chemical analyses to perform tests at different concentrations. In previously reported micro diluters, although they could provide a wide dilution range, only limited dilution ratios were demonstrated, i.e. low flexibility in dilution ratio. Here, a droplet-based micro diluter with high flexibility on dilution ratio is proposed, fabricated, and tested. Through combination of five different electrode sizes using electrowetting-on-dielectric (EWOD) technique, the proposed micro diluter can provide 49 dilution ratios at dilution range from 1/2 to 1/50. Experimental results verify that different dilution ratios can be successfully generated, even at the most difficult low concentration range. Comparing to the design with equal electrode size in conventional EWOD-based microfluidic chip, the proposed design with unequal electrode sizes can reduce the operation steps to achieve high flexibility on dilution ratios efficiently.
更多
查看译文
关键词
Droplet Size,High Flexibility,Dilution Ratio,Driving Voltage,Electrode Size