Assessing a single targeted next generation sequencing for human leukocyte antigen typing protocol for interoperability, as performed by users with variable experience.

Human immunology(2017)

引用 11|浏览57
暂无评分
摘要
BACKGROUND:A simplified protocol for HLA-typing -by NGS, developed for use with the Illumina MiSeq, was performed by technologists with varying NGS experience to assess accuracy and reproducibility. METHODS:Technologists from six laboratories typed the same 16 samples at HLA-A, B, C, DRB1, and DQB1. The protocol includes long range PCR, library preparation, and paired-end 250bp sequencing. Two indexing strategies were employed: locus-specific indexing whereby each locus was tagged uniquely and sample-specific indexing whereby all 5 loci for a sample were pooled prior to library preparation. Sequence analysis was performed with two software packages, Target HLA (Omixon) and NGSengine (GenDx). RESULTS:The average number of sequence reads per library was 387,813; however, analysis was limited to 40,000 reads for locus-indexed libraries and 200,000 reads for sample-indexed libraries resulting in an average depth of coverage of 1444 reads per locus. Sufficient reads for genotype analysis were obtained for 98.4% of libraries. Genotype accuracy was >97% in pooled amplicons and >99% in individual amplicons by both software analysis. Inter-laboratory reproducibility was 99.7% and only cause of discordance was cross-contamination of a single amplicon. CONCLUSIONS:This NGS HLA-typing protocol is simple, reproducible, scalable, highly accurate and amenable to clinical testing.