A reliable method for avoiding false negative results with Luminex single antigen beads; evidence of the prozone effect.

B Sean Carey, Kim Boswijk, Mazen Mabrok,Peter A Rowe,Andrew Connor, Imran Saif,Anthony Poles

Transplant immunology(2016)

引用 9|浏览6
暂无评分
摘要
Luminex single antigen bead (SAB) assays have become an essential tool in monitoring the status of antibody to the Human Leucocyte Antigen (HLA) of patients both before and after transplantation. In addition SAB data is used to aid risk stratification to assess immunological risk of humoral rejection in solid organ transplantation (CTAG/BTAG guidelines) [1]. Increasingly laboratories are reporting false negative results at high antibody titre due to a prozone effect. Here we report a case study where the prozone effect led to a false negative antibody result that could have resulted in adverse outcome. We describe a method to reliably remove the prozone effect through heat inactivation and the addition of Ethylenediaminetetraacetic acid (EDTA) to the Luminex wash buffer.