STIC3 – Silicon Photomultiplier Timing Chip with picosecond resolution

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2015)

引用 15|浏览21
暂无评分
摘要
The diagnostic of pancreas and prostate cancer is a challenging task due to the background noise coming from the closer organs. The EndoToFPET-US project aims to combine the synergy between metabolic and anatomical (ultrasound) image in order to improve the precision in the tumor localization. The goal of the project is to develop a Positron Emission Tomography (PET) system that provides a time-of-flight resolution of 200ps FWHM for improving the signal to noise ratio and further to improve the medical image quality. In order to achieve this purpose an ASIC has been designed for very high timing resolution in time-of-flight (ToF) applications. In this paper we present the ASIC performance and the first characterization measurements with the 64-channels prototype version (STiC3). Measurements are performed with LYSO scintillator crystal and a Multi Pixel Photon Counter (MPPC). Measurements with the chip show an analog-front-end stage jitter of 35ps for the first photo-electron equivalent charge and reach 18ps for the third photo-electron. Coincidence time resolution (CTR) of 240ps FWHM is measured with 3.1×3.1×15mm3 LYSO crystal and 50μm pixel pitch MPPC. Further optimization including the Time-to-Digital Converter (TDC) non-linearity corrections and setup fine tuning are ongoing for achieving the desired CTR of 200ps FWHM.
更多
查看译文
关键词
Time-of-flight,Front-end electronic,Positron Emission Tomography (PET)