Development of a low-temperature germanium detector via mechanical cooling with a compact pulse-tube refrigerator

T Koike,N Chiga, T Haruyama,K Hosomi, H Ichinohe,K Kasami, Takashi Matsushita, Yoshimi Mizoguchi, Hiroshi Ono,K Shirotori, Toru Takeuchi,Hirotaka Tamura,Takeshi Yamamoto

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2015)

引用 11|浏览9
暂无评分
摘要
We have developed a new germanium (Ge) detector unit for hypernuclear γ-ray spectroscopy at J-PARC, which comprises a new Ge detector array Hyperball-J. A compact pulse-tube refrigerator is coupled to a coaxial Ge detector in order to achieve lower crystal temperatures as a means to increase radiation hardness. The obtained crystal temperature is 72K, while an energy resolution is maintained at (FWHM) 3.1(1)keV for 1.33MeV γ rays using a gate-integrated shaping amplifier (ORTEC 973U). Gain shifts with changing crystal temperature are also confirmed for the Ge detector.
更多
查看译文
关键词
Germanium detector,Pulse-tube refrigerator,Gamma-ray spectroscopy,Hypernucleus