Performance of a Small Anode Germanium Well detector

A S Adekola, J Colaresi, J Douwen,W F Mueller,K M Yocum

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2015)

引用 11|浏览6
暂无评分
摘要
The performance of Small Anode Germanium (SAGe) Well detector [1] has been evaluated for a range of sample sizes and geometries counted inside the well, on the end cap or in Marinelli beakers. The SAGe Well is a new type of low capacitance germanium well detector manufactured using small anode technology. The detector has similar energy resolution performance to semi-planar detectors, and offers significant improvement over the Coaxial and existing Well detectors. Resolution performance of 0.75keV Full Width at Half Maxiumum (FWHM) at 122keV γ-ray energy and resolution of 2.0–2.3keV FWHM at 1332keV γ-ray energy are guaranteed. Such outstanding resolution performance will benefit environmental applications in revealing the detailed radionuclide content of samples, particularly at low energy, and will enhance the detection sensitivity resulting in reduced counting time. This paper reports the counting performance of SAGe Well detector for range of sample sizes and geometries and how it compares to other detector types.
更多
查看译文
关键词
Small Anode Germanium Well detector,Minimum detectable concentration,Detection sensitivity,Measurement time,Filter paper,Marinelli beaker
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要