Microneedle Electrode Array for Electrical Impedance Myography to Characterize Neurogenic Myopathy

Annals of Biomedical Engineering(2015)

引用 17|浏览4
暂无评分
摘要
Electrical impedance myography (EIM) is a noninvasive technique for neuromuscular assessment, wherein a low-intensity alternating current is applied to a muscle, and the consequent surface voltage patterns are evaluated. Commercial wet electrodes are most commonly used for EIM. However, these electrodes are not suitable for use on small muscles, as they do not effectively solve the problem of high electrode-skin contact impedance (ESCI) that negatively influences the quality of recorded biopotentials. To address this problem, we fabricated a novel microneedle electrode array (MEA) that consists of 124- µ m-long microneedles. Compared to wet electrodes, the MEA could pierce through the outer skin surface in a painless and micro-invasive manner, and could thus effectively reduce ESCI. The MEA has excellent test–retest reproducibility, with intraclass correlation coefficients exceeding 0.920. When used in combination with EIM, the MEA differentiated the affected muscles from the unaffected muscles in patients with neurogenic myopathy, by using EIM parameters of reactance and phase ( p = 0.023 and 0.008, respectively). Thus, the novel MEA is a practical and reusable device for EIM assessment in cases of neurogenic myopathy. However, further refinement of the electrode is needed to enhance the clinical application of the system.
更多
查看译文
关键词
Muscle impedance,Neuromuscular assessment,Microelectrode,Painless,Reproducibility