Performance evaluation and clinical applications of 3D plenoptic cameras

Proceedings of SPIE(2015)

引用 7|浏览16
暂无评分
摘要
The observation and 3D quantification of arbitrary scenes using optical imaging systems is challenging, but increasingly necessary in many fields. This paper provides a technical basis for the application of plenoptic cameras in medical and medical robotics applications, and rigorously evaluates camera integration and performance in the clinical setting. It discusses plenoptic camera calibration and setup, assesses plenoptic imaging in a clinically relevant context, and in the context of other quantitative imaging technologies. We report the methods used for camera calibration, precision and accuracy results in an ideal and simulated surgical setting. Afterwards, we report performance during a surgical task. Test results showed the average precision of the plenoptic camera to be 0.90mm, increasing to 1.37mm for tissue across the calibrated FOV. The ideal accuracy was 1.14mm. The camera showed submillimeter error during a simulated surgical task.
更多
查看译文
关键词
Imaging systems,Three-dimensional (3D) image acquisition,Medical and biological imaging,plenoptic camera,calibration,quantitative measurement