4 Pi Direction Sensitive Gamma Imager With Rena-3 Readout Asic - Art. No. 67060v

HARD X-RAY AND GAMMA-RAY DETECTOR PHYSICS IX(2007)

引用 7|浏览4
暂无评分
摘要
A 4 pi direction-sensitive gamma imager is presented, using a 1 cm(3) 3D CZT detector from Yinnel Tech and the RENA-3 readout ASIC from NOVA R&D. The measured readout system electronic noise is around 4-5 keV FWHM for all anode channels. The measured timing resolution between two channels within a single ASIC is around 10 ns and the resolution is 30 ns between two separate ASIC chips. After 3D material non-uniformity and charge trapping corrections, the measured single-pixel-event energy resolution is around 1% for Cs-137 at 662 keV using a 1 cm(3) CZT detector from Yinnel Tech with an 8 x 8 anode pixel array at 1.15 mm pitch. The energy resolution for two pixel events is 2.9%. A 10 uCi Cs-137 point source was moved around the detector to test the image reconstruction algorithms and demonstrate the source direction detection capability. Accurate source locations were reconstructed with around 200 two-pixel events within a total energy window 10 keV around the 662 keV full energy peak. The angular resolution FWHM at four of the five positions tested was between 0.05-0.07 steradians.
更多
查看译文
关键词
CZT detector, Compton imaging, 3D position sensing, list mode MLEM