Estimation And Removal Of The Stepped Deflation Artefact (Sda) In Oscillometric Blood Pressure Measurement

5TH EUROPEAN CONFERENCE OF THE INTERNATIONAL FEDERATION FOR MEDICAL AND BIOLOGICAL ENGINEERING, PTS 1 AND 2(2012)

引用 0|浏览3
暂无评分
摘要
A new method to detect the blood pressure pulsation signal in oscillometric blood pressure measurement using stepped deflation method was developed. In stepped deflation method, morphology of the artefact signal changes much by bleed valve switching scenario. Various types of interference caused by the bleed valve switching are analyzed. The Adaptive Impulse Correlated Filter (AICF) was applied to remove the stepped deflation artefact (SDA). This AICF is performed using a parametric method which uses Finite Impulse Response (FIR) structure. The method is implemented using a specially designed acquisition system for this research. The analysis showed that amplitude and power of artefact increased with pulse-width of the bleed valve control pulse. The removal of artefact was almost accomplished without reducing the information in the blood pressure pulsation signal.
更多
查看译文
关键词
Oscillometric blood pressure measurement, stepped deflation, Adaptive Impulse Correlated Filter (AICF)