Research On Calibration Method Of Comb Generator

2015 40TH INTERNATIONAL CONFERENCE ON INFRARED, MILLIMETER AND TERAHERTZ WAVES (IRMMW-THZ)(2015)

引用 24|浏览4
暂无评分
摘要
We describe two calibration methods of comb generator(CG) as harmonic phase reference, one is based on digital sampling oscilloscope(DSO) and the other is based on electro-optic sampling system(EOS). We calibrate Agilent U9391F CG using these two methods and compare the differences between the results. The conclusion of the research has the potential to serve the phase calibration of nonlinear instruments and devices like nonlinear vector network analyzer(NVNA), large signal network analyzer(LSNA), amplifier, etc.
更多
查看译文
关键词
comb generator,phase calibration method,CG,harmonic phase reference,digital sampling oscilloscope,electrooptic sampling system,Agilent U9391F,nonlinear vector network analyzer,large signal network analyzer,amplifier,DSO,EOS,NVNA,LSNA