Ax-Pet: Concept, Proof Of Principle And First Results With Phantoms

P. Beltrame, E. Bolle, A. Braem,C. Casella, E. Chesi, N. Clinthorne,R. De Leo,G. Dissertori, L. Djambazov,V. Fanti, C. Joram, H. Kagan, W. Lustermann,F. Meddi,E. Nappi,F. Nessi-Tedaldi,J. F. Oliver, F. Pauss,M. Rafecas, D. Renker, A. Rudge, U. Ruotsalainen,D. Schinzel,T. Schneider,J. Seguinot,P. Solevi, S. Stapnes,U. Tuna, P. Weilhammer

2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC)(2010)

引用 4|浏览12
暂无评分
摘要
AX-PET is a novel PET concept based on long crystals axially arranged and orthogonal Wavelength shifter (WLS) strips, both individually readout by Geiger-mode Avalanche Photo Diodes (G-APD). Its design was conceived in order to reduce the parallax error and simultaneously improve spatial resolution and sensitivity. The assessment of the AX-PET concept and potential was carried out through a set of measurements comprising individual module characterizations and scans in coincidence mode of point-like and extended sources. The estimated energy and spatial resolutions from point-like measurements are R-FWHM = 11.6% (at 511 keV) and 1.7-1.9 mm (FWHM) respectively as measured with point-like sources placed in different positions of the FOV. First results from scans of extended phantoms confirmed our expectations.