Temperature Monitoring By Infrared Radiation Measurements During Arf Excimer Laser Ablation With Cornea

M Ishihara,T Arai,S Sato,H Nakano,M Obara, M Kikuchi

LASER-TISSUE INTERACTION X: PHOTOCHEMICAL, PHOTOTHERMAL, AND PHOTOMECHANICAL, PROCEEDINGS OF(1999)

引用 3|浏览2
暂无评分
摘要
We measured infrared thermal radiation from porcine cornea during various fluences ArF excimer laser ablations with 1 mu s rise time. To obtain absolute temperature by means of Stefan-Boltzman law of radiation, we carried out a collection efficiency and detective sensitivity by a pre-experiment using panel heater. We measured the time course of the thermal radiation intensity with various laser fluences. We studied the relation between the peak cornea temperature during the ablation and irradiation fluences. We found the ablation situations, i.e., sub-ablation threshold, normal thermal ablation, and over-heated ablation, may be judged by both of the measured temperature transient waveformes and peak temperature. The boundary fluences corresponding to normal thermal ablation were 90 and 160mJ/cm2. Our fast remote temperature monitoring during cornea ablation might be useful to control ablation quality/quantity of the cornea ArF laser ablation, that is PRK.
更多
查看译文
关键词
ArF excimer laser,cornea ablation,temperature,thermal radiation,remote monitoring,thermal radiation,Photorefractive Keratectomy,PRK,Stefan-Boltzman law of radiation