Evaluation of MALDI-ToF as a method for the identification of bacteria in the veterinary diagnostic laboratory

L P Randall, F Lemma,Mark S Koylass,Jon Rogers,R D Ayling, Danny Worth, Monika Klita, Andrew Steventon, Kirsty Line, Peter D Wragg,Jakub Muchowski,Markus Kostrzewa,A M Whatmore

Research in Veterinary Science(2015)

引用 68|浏览22
暂无评分
摘要
Matrix-Assisted Laser Desorption Ionisation-Time of Flight (MALDI-ToF) Mass Spectrometry with Bruker MALDI Biotyper software was evaluated as a method for identifying veterinary bacteria. For 620 isolates (~100 bacterial species), identification by MALDI-ToF and non-16S rDNA methods (mainly phenotypic/biochemical) agreed to species-level (95.3%) and to species/genus-level (100%), but in the absence of 16S rDNA as a gold standard. For a further panel of 107 anaerobes and 234 aerobes (~100 bacteria species) using 16S rDNA results as the gold standard, MALDI-ToF/biochemical tests showed 97.8/96.6% species-level and 99.6/93.5% genus-level agreement for aerobes and 95.3/93.6% species-level and 100/95.3% genus-level agreement for anaerobes compared to the gold standard. Where results were obtained from direct spots, direct spots overlaid with formic acid and extracts, 89.4% of 180 aerobes and 90.1% of 152 anaerobes were identified by MALDI-ToF. MALDI-ToF was shown to be a rapid and reliable method to identify veterinary bacteria.
更多
查看译文
关键词
Bacterial identification,16S rDNA,Biochemical,MALDI-ToF