Diagnostic system of Hα emission on neutral beam injection experiments in the GAMMA10 central cell

REVIEW OF SCIENTIFIC INSTRUMENTS(2004)

引用 1|浏览14
暂无评分
摘要
We newly designed and installed a Halpha line-emission detector in order to examine neutral particle transport around the beam line of NBI in the central cell of the GAMMA10 tandem mirror. The Halpha detector consists of a Halpha interference filter, a lens, an optical fiber, and a photomultiplier. The detector is absolutely calibrated using a standard lamp. The Halpha detector is located at the angle of 24degrees to the beam line, which enabled us to directly view the Halpha emission due to neutral beam injection (NBI). In standard hot-ion mode plasmas, we measured the Halpha line emission in the case of NBI. The Halpha line emission during NBI strongly increased by 550% on the beam line compared with that before NBI. Moreover, it was found that hydrogen recycling played an important role in the neutral transport around the beam line. This diagnostic system is confirmed to be considerably helpful for examining the detailed behavior of the neutral particles related to NBI for the first time. (C) 2004 American Institute of Physics.
更多
查看译文
关键词
optical fiber,gas dynamics
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper