First characterization of the SPADnet sensor: a digital silicon photomultiplier for PET applications

JOURNAL OF INSTRUMENTATION(2013)

引用 10|浏览22
暂无评分
摘要
Silicon Photomultipliers have the ability to replace photomultiplier tubes when used as light sensors in scintillation gamma-ray detectors. Their timing properties, compactness, and magnetic field compatibility make them interesting for use in Time-of-Flight Magnetic Resonance Imaging compatible Positron Emission Tomography. In this paper, we present a new fully digital Single Photon Avalanche Diode (SPAD) based detector fabricated in CMOS image sensor technology. It contains 16x8 pixels with a pitch of 610x571.2 mu m(2). The Dark Count Rate and the Photon Detection Probability of each SPAD has been measured and the homogeneity of these parameters in the entire 92000 SPAD array is shown. The sensor has been optically coupled to a single LYSO needle and a LYSO array. The scintillator crystal was irradiated with several gamma sources and the resulting images and energy spectra are presented.
更多
查看译文
关键词
Gamma detectors (scintillators, CZT, HPG, HgI etc),Photon detectors for UV, visible and IR photons (solid-state),Gamma camera, SPECT, PET PET/CT,coronary CT angiography (CTA)