Determination of urinary coenzyme Q10 by HPLC with electrochemical detection: Reference values for a paediatric population.

BIOFACTORS(2015)

引用 20|浏览12
暂无评分
摘要
Kidney dysfunction is being increasingly associated with mitochondrial diseases and coenzyme Q(10) (CoQ) deficiency. The assessment of CoQ status requires the biochemical determination of CoQ in biological fluids and different cell types, but no methods have been developed as yet for the analysis of CoQ in excretory systems. The aim of this study was to standardize a new procedure for urinary CoQ determination and to establish reference values for a paediatric population. Urinary CoQ was analyzed by HPLC with electrochemical detection. Reference values (n=43) were stratified into two age groups (2-10 years: range 24-109 nmol CoQ/gram of pellet protein; 11-17 years: range 43-139 nmol CoQ/gram of pellet protein). In conclusion, urinary CoQ analysis is a noninvasive, reliable, and reproducible method to determine urinary tract CoQ status. (c) 2015 BioFactors, 41(6):424-430, 2015
更多
查看译文
关键词
coenzyme Q10,mitochondrial disease,renal disease,urine,high pressure liquid chromatography,reference values