Imaging Of Cardiac Electrical Excitation Conduction

D. F. Zhou,S. Q. Jiang, J. C. Zhu, C. Zhao, Y. R. Yan, D. Groenemeyer,P. Van Leeuwen

2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC)(2015)

引用 1|浏览0
暂无评分
摘要
We present a multiple time windows beamformer (MTWB) method of solving the inverse problem of magnetic field and non-invasively imaging the cardiac electrical excitation conduction using the magnetocardiac signals acquired by a 61-channel superconducting quantum interference device (SQUID). The MTWB constructs spatial filters for each location in source space, one for each component of the source moment based on the distributed source model, and estimates the cardiac equivalent current sources. The output of spatial filters is the source strength estimated in three-dimensional space and the weight matrix calculated with magnetocardiac signals in multiple time windows. A signal subspace projection technique is used to suppress noise. Then, the characteristics of cardiac electrical excitation conduction among two healthy subjects and two coronary vessel stenosis (CVS) patients are extracted from reconstructed current sources with maximum strength at each instant during QRS complex and ST-T segment, and a series of two-dimensional cardiac electrical excitation conduction maps (EECM) are obtained. It is demonstrated that two healthy subjects are of similar and the stronger electrical activities than those of two CVS patients. This technique can be used as an effective tool for the diagnosis of heart diseases.
更多
查看译文
关键词
Heart,Heart Rate,Humans