Drilling the near cortex with elongated figure-of-8 holes to reduce the stiffness of a locking compression plate construct.

JOURNAL OF ORTHOPAEDIC SURGERY(2015)

引用 2|浏览22
暂无评分
摘要
Purpose. To compare the stiffness of locking compression plate (LCP) constructs with or without drilling the near cortex with elongated figure-of-8 holes. Methods. 24 synthetic bones were sawn to create a 10-mm gap and were fixed with a 9-hole 4.5-mm narrow LCP. In 12 bones, the near cortex of the adjacent holes to the LCP holes was drilled to create elongated figure-of-8 holes before screw insertion. The stiffness of LCP constructs under axial loading or 4-point bending was assessed by (1) dynamic quasi-physiological testing for fatigue strength, (2) quasi-static testing for stiffness, and (3) testing for absolute strength to failure. Results. None of the 24 constructs had subcatastrophic or catastrophic failure after 10 000 cycles of fatigue loading (p=1.000). The axial stiffness reduced by 16% from 613 +/- 62 to 517 +/- 44 N/mm (p=0.012) in the case group, whereas the bending stiffness was 16 +/- 1 Nm(2) in both groups (p=1.000). The maximum axial load to catastrophic failure was 1596 +/- 84 N for the control group and 1627 +/- 48 N for the case group (p=0.486), whereas the maximum bending moment to catastrophic failure was 79 +/- 12 and 80 +/- 10 Nm, respectively (p=0.919). Conclusion. Drilling the near cortex with elongated figure-of-8 holes reduces the axial stiffness of the LCP construct, without compromising its bending stiffness or strength.
更多
查看译文
关键词
bone plates,compressive strength,fracture fixation