Use of the Off-the-Shelf t-Branch Device to Treat an Acute Type Ia Endoleak in a Symptomatic Juxtarenal Abdominal Aortic Aneurysm.

JOURNAL OF ENDOVASCULAR THERAPY(2016)

引用 9|浏览3
暂无评分
摘要
Purpose: To report the use of the off-the-shelf Zenith t-Branch Thoracoabdominal Endovascular Graft for the treatment of a symptomatic acute type Ia endoleak subsequent to previous infrarenal bifurcated endovascular aneurysm repair. Case Report: A 72-year-old man presented 4 years after implantation of an Anaconda stent-graft with sac expansion and type II endoleak, which was treated over the next 18 months with repeated coil and glue embolization of the leak and sac. Six months after the last attempt, imaging disclosed a new type Ia endoleak. Sac enlargement became symptomatic, and emergent treatment was performed using the off-the-shelf Zenith b-Branch device. The 4 visceral arteries were bridged with Fluency stent-grafts. The Anaconda stent-graft precluded the use of a standard Zenith universal distal body bifurcated device below the t-Branch graft, so a reversed tapered 16x20x82-mm iliac limb was placed to bridge the t-Branch to the larger Anaconda limb; the smaller 12-mm contralateral Anaconda limb was occluded. The procedure was concluded with a femorofemoral bypass. At 10 months after the procedure, the patient is well; the sac diameter has decreased, and there is no endoleak on imaging. Conclusion: The multibranched Zenith t-branch device may be useful in treating proximal endoleaks in bifurcated stent-grafts.
更多
查看译文
关键词
abdominal aortic aneurysm,bifurcated stent-graft,endovascular aneurysm repair,juxtarenal aneurysm,multibranched stent-graft,sac enlargement,t-branch device,type la endoleak
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要