Analytical performance of the Alere™ i Influenza A&B assay for the rapid detection of influenza viruses.

Enfermedades Infecciosas y Microbiología Clínica(2017)

引用 11|浏览3
暂无评分
摘要
The analytical performance of the new Alere™ i Influenza A&B kit (AL-Flu) assay, based on isothermal nucleic acids amplification, was evaluated and compared with an antigen detection method, SD Bioline Influenza Virus Antigen Test (SDB), and an automated real-time RT-PCR, Simplexa™ Flu A/B & VRS Direct assay (SPX), for detection of influenza viruses. An “in-house” RT-PCR was used as the reference method. Sensitivity of AL-Flu, SDB, and SPX was 71.7%, 34.8%, and 100%, respectively. Specificity was 100% for all techniques. The turnaround time was 13min for AL-Flu, 15min for SDB, and 75min for SPX.
更多
查看译文
关键词
Influenza,Rapid detection,Molecular testing