Wire In The Heart: Fracture And Fragment Embolization Of Retrievable Inferior Vena Cava Filter Into The Right Ventricle

CASE REPORTS IN CARDIOLOGY(2015)

引用 5|浏览1
暂无评分
摘要
We report a case of a 58-year-old female who was found to have a fractured limb of her IVC filter in her right ventricle during a cardiac catheterization. A 25 mm radioopaque thin linear structure was seen in the proximal portion of the right ventricle. It was fixed and did not migrate or change position during investigations. On fluoroscopy, the IVC filter was observed in an appropriate location in the midabdomen. Yet, fractures of at least two of the metal filamentous legs of the IVC device were noticed. The patient was made aware of the many risks associated with filter removal. Due to the high risks of the procedure, she refused surgery and the filter fragment was not removed. We present this case to underscore the potential complications of IVC filters.